FDA Device

Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

Hazard

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

What You Should Do

Remedy

On 04/07/2021, Roche issued Urgent Medical Device Correction via fax notifying customers that Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/ proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte was measured immediately before the Troponin I analyte on the same module, a decreased recovery of Troponin I was observed.

Products

Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
Brand
Roche Diagnostics Operations, Inc.
UPC
UDI: 04015630939381

Units Affected

10, 489 kits