FDA Device

St. Jude Medical Confirm Rx Insertable Cardiac Monitor, REF DM3500, Sterile. Product Usage: The St. Jude Medical Confirm Rx insertable cardiac monitor (ICM) is designed to detect arrhythmias and wirelessly transmit data to the Merlin.net Patient Care Network (PCN). It is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for card

Hazard

The device is unable to pair with the mobile app due to the device incorrectly determining the certificate has expired.

What You Should Do

Remedy

Abbott sent an Important Medical Device Correction letter dated November 20, 2018 to affected customers. The letter identified the affected product, problem and actions to be taken. Should you have questions about this issue, contact your local Abbott Representative or Abbott Support at 1-800-727-7846 (Opt3) (U.S.), 8:30am - 5:30pm Central Time Monday thru Friday.

Products

St. Jude Medical Confirm Rx Insertable Cardiac Monitor, REF DM3500, Sterile. Product Usage: The St. Jude Medical Confirm Rx insertable cardiac monitor (ICM) is designed to detect arrhythmias and wire
Brand
Abbott
UPC
Serial numbers 9225191, 9224501, 9225179, 9225193, 9225207, 9242209, 9257638, 9257646, 9257656, 9257659, 9224508, 9224490, 9224503, 9224525, 9224529, 9225190, 9225203, 9225205, 9225208, 9225211, 9225214, 9225851, 9225855, 9225858, 9225859, 9225867, 9225868, 9242199, 9242219, 9242225, 9242231, and 9256057.

Units Affected

32 devices