CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS
Hazard
There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
What You Should Do
Remedy
The firm issued a field safety notice by letter on 04/10/2019. The letter explained the issue and provided safety instructions. All impacted customers with affected systems will be contacted to arrange for a software upgrade.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
CyberKnife M6, Part Numbers:
a) 0660000: CyberKnife Treatment Delivery System
b) 053101-010: CyberKnife M6 FI Treatment Delivery System
c) 053201-010: CyberKnife M6 FM Treatment Delivery System
d) 0
- Brand
- Accuray Incorporated
- UPC
- a) 0660000: CyberKnife Treatment Delivery System Serial Numbers: C0434, C0436, C0437, C0438, C0440, C0441, C0443, C0444, C0445, C0446, C0447, C0448, C0449, C0450, C0453, C0454, C0456, C0457, C0458, C0460, C0464, C0465, C0466, C0472, C0475 b) 053101-010: CyberKnife M6 FI Treatment Delivery System Serial Numbers: C0334, C0381 c) 053201-010: CyberKnife M6 FM Treatment Delivery System Serial Numbers: C0363 d) 053301-010: CyberKnife M6 FIM Treatment Delivery System Serial Numbers: C0337, C0343, C0349, C0352, C0358, C0370, C0371, C0378, C0384, C0387, C0389, C0390, C0391, C0393, C0409, C0424, C0426, C0427, C0428, C0433, C0439 e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System Serial Numbers: C0330, C0332, C0342, C0344, C0345, C0348, C0359, C0360, C0364, C0367, C0372, C0394, C0398, C0400, C0402, C0404, C0415, C0419 f) 053501-010: CyberKnife Matrix Tier F TDS Serial Numbers: C0396
Units Affected
68 units