GE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

Hazard

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

What You Should Do

Remedy

The firm initiated their action by letter to the consignees on 12/10/2020. The letter explained the issue and stated that GE would inspect, and if required, correct the affected units. Questions and concerns may be directed to GE Healthcare Service at 1-800-437-1171 or the consignee's local Service Representative.

Products

GE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
Brand
GE Healthcare, LLC
UPC
Serial numbers: 85EY67048, 850Y61018, 85EY67045, 85EY67046, 85EY67047, 850Z61038, 85EY67043, 850Z61042, 85EZ67056, 850Z61043,  850Z61044, 850Z61045, 850Z61041, 850Z61039, 850Z61040, 85EZ67055, 85EZ67053, 85EZ67054, 850Y61019, 85EY67044, 85EY67049, 85EZ67052, 85EY67050, 85EZ67051

Units Affected

24 units