FDA Device

Yukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (the Alignment Tube ) provides counter rotation to reduce the torque applied to the anatomy when the size 15 Torque limiting shaft (the shaft is used to tighten the Yukon Set Screw to the specified torque.

Hazard

Shafts do not pass through the inner cannula of the Anti-Torque Alignment Tube from Lot KTGG due to a manufacturing nonconformance. This could result in a delay of surgery or result in surgical intervention later on.

What You Should Do

Remedy

Our records indicate Stryker previously distributed catalog no. 7601-9001, Lot KTGG to your location in a Stryker Loaner Set. The nonconforming instrument was subsequently identified and quarantined during routine processing of loaner sets by Stryker at its Leesburg facility. This communication is being made for notification purposes only. The only action requested of you is to complete the Business Reply form found on the following page and return it to the email addresses referenced in the form. If you have any questions or concerns, please contact Stryker Regulatory Compliance at 201-749-8090. On behalf of Stryker we thank you sincerely for your help and support in completing the Business Reply Form and returning it to Stryker by November 6th, 2020.

Products

Yukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (the Alignment Tube ) provides counter rotation to reduce the torque applied to the anatomy wh
UPC
Catalog Number/REF 7601-90001 LOT KTGG GTIN 10888857343948

Units Affected

4 devices