FDA Device

SMARTABLATE RF System Generator

Hazard

Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets.

What You Should Do

Remedy

All new U.S. and OUS customers, who have purchased the SMARTABLATE" Generator system since the launch of the original field action in March 2015 to the time the IFU was updated in November 2015, will be notified. The previously distributed Field Safety Notification Letter for the SMARTABLATE" Generator (Part Numbers# M490006 and M490007) and customer Acknowledgement Form will be sent to these new customers. Instructions included review the Field Safety Notice carefully, provide the notice to appropriate personnel in facility, ensure the SMARTABLATE System Foot Pedal is not located within the Stereotaxis RMT Lab, complete and return the Acknowledgement Forms, and maintain awareness of the notice.

Products

SMARTABLATE RF System Generator
Brand
Biosense Webster, Inc.
UPC
Product Codes M490006, M490007

Units Affected

185 units total