FDA Device

CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.

Hazard

Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.

What You Should Do

Remedy

The firm initiated their market withdrawal by letter on 08/28/2018. The firm requested surgeons immediately cease implantation of the device and return any unused product.

Products

CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.
Brand
Alcon Research, LTD.
UPC
GTIN: 00380657540037 All lot/serial numbers

Units Affected

30561 units