Medivators Endo Smartcap Endoscope Tubing. REF100145. Packaged as Contents 1, 10, 50, and found in BMP-C145. Tubing for Olympus 140/240, 160/260, 180/280, 190/290 Series Endoscopes. Sterile EO - Product Usage: is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.
Hazard
Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.
What You Should Do
An Urgent Medical Device Recall Notification letter was sent to customers via certified mail next day delivery on 10/09/2019. The letter identified affected product, stated reason for recall, asked for product to be quarantined and discontinue use. Customers are to return product to Medivators by obtaining a Return Material Authorization from Customer Service. A response form was asked to be returned. Questions can be directed to Customer Service at 1-800-444-4729.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Medivators, Inc.
- UPC
- Lot # / UDI #s: 427416 / 00677964059492208051427416, 427098 / 00677964059492208031427098, 427819 / 00677964059492208071427819, 427154 / 00677964059492208031427154.
Units Affected
53,790 units total