MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients
Hazard
The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.
What You Should Do
Remedy
This screen would not be accessed by normal user (only servicer). Firm is notifying servicing personnel and not consumer/user level. No customer letters/communication sent. Servicer will install edge guard on effected units. Technical Service Bulletin issued to servicer.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients
- Brand
- Mevion Medical Systems, Inc.
- UPC
- Lot #s: MEVION S250-0001 MEVION S250-0002 MEVION S250-0003 MEVION S250-0004 MEVION S250-0005 MEVION S250-0006 MEVION S250-0007 MEVION S250-0008 UDI: (01)00864366000100 (11) 131209 (01)00864366000100 (11)150526 (01)00864366000100 (11)150113 (01)00864366000100 (11)150430 (01)00864366000100 (11)160127 (01)00864366000100 (11)160603 (01)00864366000124 (11) 180220 (01)00864366000124 (11) 181219
Units Affected
8 units