BD Microtainer Tube w/ BD Microgard Closure, K2EDTA additive, Model Numbers 365974 and 365975 Product Usage: The Microtainer Brand Tube with EDTA(K2) Reorder No. 365974, is used to collect, anticoagulate, transport and store skin puncture blood specimens for testing the following hematological parameters: WBC, RBC ,HGB, HCT, MCV, MCH, MCHC, Platelets, 5 part differential including : Neutrophils, Lymphocytes, Monocytes, Eosinophils, Basophils, Manual WBC Differentials, Reticulocytes, and Spun He
Hazard
The products have been confirmed to have reduced or no additive within the tube reservoir. The tubes may develop visible clots within the tube samples or micro clots that are not easily detected during visual inspection of the tubes. A clotted sample may lead to recollection of samples or, retesting of patients, resulting in delayed reporting of test results and patient treatment.
What You Should Do
Remedy
Urgent Medical Device Recall notification letters dated 6/3/19 were sent to customers. Thee letter identified the affected product, problem and actions to be taken.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
BD Microtainer Tube w/ BD Microgard Closure, K2EDTA additive, Model Numbers 365974 and 365975
Product Usage:
The Microtainer Brand Tube with EDTA(K2) Reorder No. 365974, is used to collect, anticoagu
- Brand
- Becton Dickinson & Company
- UPC
- Lot Numbers: 9017588 9017590 9017591 9017592 9017593 9017596 9017600 9028803 9045873 9046822 9050660 9052823 9056896
Units Affected
1707804