FDA Device

VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents

Hazard

One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.

What You Should Do

Remedy

Urgent Product Correction Notification letters dated 3/11/19 were sent to customers.

Products

VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and
Brand
Ortho-Clinical Diagnostics
UPC
Systems Running Software Version 3.4 & 3.4.1

Units Affected

80