FDA Device

TLC-2000 Therapeutic Medical Laser System, used for the temporary relief of knee pain.

Hazard

Device was shipped without a Unique Device Identification (UDI) label.

What You Should Do

Remedy

On September 26, 2019 the firm distributed Urgent Medical Device Recall Communication letters. Check the Serial Number on your TLC-2000 Power Pack (Model Number: TLC-2002) to see if your TLC-2000 is affected by the following issues: 1) The TLC-2000 Power Pack was shipped with a label identifying it as being CE compliant. What Should You Do?: No action required. The TLC-2000 is approved for sale in Canada under Medical Device License Number: 98357. The TLC-2000 is cleared for sale in the United States by the FDA under 510(k) number: K151816. *Product with Serial Number 3204 was sold in the European Union. Only the owner of Serial Number 3204 should return their TLC-2000. 2)The TLC-2000 was shipped without a Unique Device Identification (UDI) label. What Should You Do?: Please courier your TLC-2000 to Theralase, at your cost, for update. A UDI label will be added to your TLC-2000, at no charge and your TLC-2000 will be couriered back to you at Theralase's cost.

Products

TLC-2000 Therapeutic Medical Laser System, used for the temporary relief of knee pain.
Brand
Theralase Inc.
UPC
Serial Numbers: 3015, 3022, 3037, 3040, 3047, 3052, 3055, 3074, 3076, 3080, 3100, 3101, 3102, 3103, 3123, 3127, 3129, 3131, 3132, 3137, 3139, 3155, 3162, 3166, 3171, 3183, 3187, 3200, 3206, 3227, 3245, 3266, 3268, 3293, 3296, 3297, 3305, 3308, 3318, 3321, 3332

Units Affected

41 units