TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary relief of knee pain.
Hazard
Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)
What You Should Do
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019. Healthcare practitioners are instructed to courier their TLC-2000 to Theralase. A Field Inspection Test will be completed and once passed, a Special Inspections for Medical Electrical Equipment and Systems label will be applied to the TLC-2000.The Field Inspection Test will be completed according to either CSA 60601-1 or UL 60601-1 dependent upon the customers Canadian or United States location, respectively.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Theralase Inc.
- UPC
- Serial Numbers: 3275 to 3350 (inclusive)
Units Affected
76 units