Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.
Hazard
The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.
What You Should Do
Remedy
The recalling firm issued letters on 5/9/2019 to the consignees who had the installed instruments with the affected serial numbers. The letters were issued via FedEx priority overnight express.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.
- Brand
- Abbott Laboratories
- UPC
- Tubing lot #2018-10-16; Instrument serial numbers 007689, 007690, 007691, 007692, 007705, 007706, 007707, 007708, 007709, 007728, 007694, 007695, 007697, 007766, 007767, 007768, 007765, 007710, 007699, 007700, 007701, and 007703.
Units Affected
22 analyzers with tubing