FDA Device

The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.

Hazard

Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.

What You Should Do

Remedy

Urgent Medical Device Correction notification letters dated 6/3/19 were sent to customers.

Products

The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.
Brand
Roche Diagnostics Corporation
UPC
Lot 368756

Units Affected

8 units