FDA Device

VITROS Chemistry Products Performance Verifier is used as an assayed control to monitor the performance on VITROS 250/350/5, 1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Used as a control for the VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Hazard

VITROS Performance Verifier II (PVII) Lot K4852 does not meet long term stability requirements for the AST Analyte (only) when the product is stored refrigerated for up to 6 months.

What You Should Do

Remedy

On July 10, 2017 the firm sent notification letters and an e-communications to their customers. Foreign customers were notified by email. Customers were instructed to complete a response form with the amount of product on hand and to discontinue use of the recalled product. For further questions, please call (585) 453-3452.

Products

VITROS Chemistry Products Performance Verifier is used as an assayed control to monitor the performance on VITROS 250/350/5, 1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Us
Brand
Ortho Clinical Diagnostics Inc
UPC
Lot K4852, Expiration 22 Feb 2018

Units Affected

10,021 units