ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.
Hazard
Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).
What You Should Do
Remedy
All 9 domestic units were corrected by the firm in February 2017, via technician visits to consignee sites.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.
- Brand
- Zimmer Biomet, Inc.
- UPC
- Model No. 2.5.8., Serial No. RO10009, RO10011, RO10014, RO13023, RO13027, RO14031, RO14033, RO14035, and RO14037.
Units Affected
31 units