ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.

Hazard

Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller.

What You Should Do

Remedy

This is a retrospective report of a correction initiated on 20-Jul-2016. Customers were informed onsite by field Service Technicians of the planned correction. The software issue described was corrected in the new software version ROSA Brain 3.0.0.20. Field Service Technicians were deployed to the customers locations to perform the system upgrade. .

Products

ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments
Brand
Zimmer Biomet, Inc.
UPC
Serial #'s: BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, and  BR16021

Units Affected

18 (14 US and 4 OUS)