Arthrex Shoulder Suspension Tower; Catalog Number: AR-1650-01 Product Usage: The Arthrex Lateral Decubitus Shoulder Positioner is intended to allow for the controlled manipulation of the arm and shoulder joint during arthroscopic surgical procedures.
Hazard
During use, there is a potential scenario which could allow the fastening screws attaching covers to come loose and potentially land in the sterile field causing a delay in surgery.
What You Should Do
Remedy
The distributor was notified of the recall via email on January 3, 2017. The letter identified the affected product, problem and actions to be taken. The letter instructs the consignee to remove the product from service and follow instructions to return the device to Allen Medical for repair.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Arthrex Shoulder Suspension Tower; Catalog Number: AR-1650-01
Product Usage:
The Arthrex Lateral Decubitus Shoulder Positioner is intended to allow for the controlled manipulation of the arm and shou
- Brand
- Allen Medical Systems, Inc. One Post Office Square
- UPC
- Serial Numbers: 501798, 501799, 501800, 501801, 501802, 501803, 501804, 501805, 501806, 501807, 501808, 501809, 501810, 501811, 501812, 501813, 501814, 501815, 501816, 501817, 501818, 501819, 501820, 501821, 501822, 501823, 501824, 501825, 501826, 501827, 501828, 501829, 501830, 501831, 501832, 501833, 501834, 501835, 501836, 501837, 501838, 501839, 501840, 501841, 501842, 501848, 501849, 501850, 501851, 501852, 501853, 501854, 501855, 501856, 501857, 501863, 501864, 501865, 501866, 501867, 501868, 501869, 501870, 501871, 501872, 501878, 501879, 501880, 501881, 501882, 501883, 501884, 501885, 501886, 501887, 501888, 501889, 501890, 501891, 501892, 501893, 501894, 501895, 501896, 501897, 510733, 510734, 510735, 510736, 510737, 510738, 510739, 510740, 510741, 510742, 510743, 510744, 510745, 510746, 510747, 510748, 510749, 510750, 510751, 510752, 510834, 510835, 510836, 510837, 510838, 510839, 510840
Units Affected
112 devices