FDA Device

Tryptic Soy Agar, Acumedia PN 7100 500g, 2Kg, 10Kg, and 50Kg sizes

Hazard

light precipitate may form in prepared plates of Acumedia Tryptic Soy Agar (7100) Lot# 108166. While there is no change in growth performance of this medium, the formation of particulate matter may be misinterpreted as growth in poured plates thereby leading to inaccurate counts.

What You Should Do

Remedy

Neogen sent a Recall Notification dated March 8, 2016, to all affected customers on April 22, 2016. via e-mail. The recall notification included a description of the reason for the recall, affected product, and offer for replacement. If any Tryptic Soy Agar from this lot remains in your inventory, and you would like to replace this lot, please contact the customer service team at Neogen Corporation by email at foodsafety@neogen.com or by telephone at (800)-234-5333. We would like to assure you that we are committed to providing products that meet our strict quality standards for your testing needs. Neogen will investigate this isolated incident to determine the root cause and implement measures to ensure that this issue does not happen in the future.

Products

Tryptic Soy Agar, Acumedia PN 7100 500g, 2Kg, 10Kg, and 50Kg sizes
Brand
Acumedia Manufacturers, Inc.
UPC
7100A Lot 108166C 7100B, Lot 108166B 7100C, Lot 108166A 7100D, Lot 108166A 7100CAR, lot 108166B 108-1, Lot 108166B

Units Affected

482