Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified medical personnel.
Hazard
The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices.
What You Should Do
On April 27, 2018, the recalling firm contacted its consignees by telephone prior to sending out Urgent Medical Device Recall notifications via e-mail. Customers were instructed to remove affected devices from operation immediately. The firm provided replacement devices so customers could send their affected ventilator to the firm's repair center for update once the required parts were available. Customers were also asked to inform all affected users at each site. If you have any questions regarding this letter please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437. For questions regarding the operation and/or servicing of your Carina Ventilator please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Draeger Medical, Inc.
- UPC
- Serial nos. ASKM-0065 ASKM-0066 ASKM-0106 ASKM-0134 ASKM-0137 ASKM-0150 ASKM-0151 ASKN-0009
Units Affected
8 units