Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized Item Number: 6000689. Supplied for use with JANUS Automated Workstations: Product Number: AJS4001 AJM4001 AJI4001, AJL4001, AJS8001 , AJM8001, AJI8001, AJL8001, AJM4G01 , AJI4G01 , AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01 , AJL4M01, AJM8M01, AJI8M01, AJL8M01, AGS4NGS The JANUS Automated Workstation is an automated, programmable liquid handling instrument intended for use in clinical settings to facilitat
Hazard
Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing
What You Should Do
PerkinElmer initiated recall on June 24, 2016 to the Product Manager and was delivered to the customer by email facilitated through PerkinElmer Sales Team.Customers were requested to take the following actions until PerkinElmer's corrective measures are completed: Inspect customer inventory of the 6000689 25 Clear Filter Pipette Tip and identify if the affected lots are present. Complete and return the attached Response Form with the required information. Destroy all quantities of tips from the affected lots. Upon receipt of the return Response Form, a shipment of replacement tips will be shipped.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- PerkinElmer, Inc
- UPC
- Lot Numbers:15470394, 15471775, 1601067, 16071147
Units Affected
595 trays (571,200 tips)