Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017
Hazard
Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.
What You Should Do
Remedy
Consignees will be notified on 5/15/17 via certified mail to return products.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017
- Brand
- Zimmer Biomet Spine Inc.
- UPC
- Model No: 07.02066.001 (All lots); 07.02053.001 (All lots distributed prior to April 26, 2017), and 730M0017 (Lot 309622)
Units Affected
497 units