VS200 radiographic Cassette Holder or Wall Stand. For use in radiology department.
Hazard
The wall stand cassette holder can suddenly fall to the bottom of its travel, near the ground and contact patient.
What You Should Do
The firm sent an email followed by a certified undated letter entitled Safety Related Notification- Field Corrective Action VS-200 and VS-300 Wallstands notifying customers of the recall and actions to be taken. To prevent the possible interruption of service, Del Medical has prepared a service kit which must be installed at the next available opportunity. Del Medical is sending service kits to affected customers. Included within the package will be needed components and, a fixture which will facilitate a one person job. The firm expects the procedure to occupy one hour or, less. Also included with each kit will be an acknowledgement form which is intended to document the fix. The form, along with the original parts are to be returned to Del Medical upon completion of the activity to the following address: Del Medical 241 Covington Drive Bloomingdale, Illinois 60108 Attention: Bill Kostecki 1-847-288-7000
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Del Medical, Inc.
- UPC
- Product # Serial # 110-5010G1 , 00504-0214 110-5010G1 , 00855-0414 110-5010G1 , 00182-0114 110-5010G1 , 00108-0114 110-5010G1 , 00479-0214 110-5010G1 , 00014-0114 110-5010G1 , 00251-0214 110-5010G1 , 00854-0414 110-5010G1 , 00505-0214 110-5010G1 , 00342-0214 110-5010G1 , 00477-0214 110-5010G1 , 00184-0114 110-5010G1 , 00248-0214 110-5010G1 , 00588-0314 110-5010G1 , 00852-0414 110-5010G1 , 00107-0114 110-5010G1 , 00183-0114 110-5010G1 , 00480-0214 110-5010G1 , 00503-0214 110-5010G1 , 00443-0214 110-5010G1 , 00853-0414 110-5010G1 , 00941-0414 110-5010G1 , 00013-0114 110-5010G1 , 00343-0214 110-5010G1 , 00478-0214 110-5010G1 , 00249-0214 110-5010G1 , 00185-0114 110-5010G1 , 00442-0214 110-5010G1 , 00632-0314 110-5010G1 , 00475-0214 110-5010G1 , 00507-0214 110-5010G1 , 00015-0114 110-5010G1 , 00631-0314 110-5010G1 , 01008-0514 110-5010G1 , 00635-0314 110-5010G1 , 00011-0114 110-5010G1 , 00012-0114 110-5010G1 , 00345-0214 110-5010G1 , 00590-0314 110-5010G1 , 00440-0214 110-5010G1 , 00591-0314 110-5010G1 , 00506-0214 110-5010G1 , 00106-0114 110-5010G1 , 00774-0314 110-5010G1 , 00508-0214 110-5010G1 , 00476-0214 110-5010G1 , 00186-0114 110-5010G1 , 00587-0314 110-5010G1 , 00441-0214 110-5010G1 , 00589-0314 110-5010G1 , 00250-0214 110-5010G1 , 00247-0214 110-5010G1 , 00773-0314 110-5010G1 , 00344-0214 110-5010G3 , 01090-0514 110-5010G3 , 00709-0314 110-5010G3 , 00988-0414 110-5010G3 , 00010-0114 110-5010G4 , 01080-0514 110-5010G4 , 00047-0114 110-5010G4 , 00925-0414
Units Affected
4000