Magnesium on RX Imola analyser IVD

Hazard

According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when setting up the magnesium assay. A correction was made to the IFU for the Magnesium assay and all RX Imola customers in the USA were contacted with the updated instruction.

What You Should Do

Remedy

Randox sent an Urgent Field Safety Notice to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to review and update Rx Imola running order where required. Rework any remaining stock with updated IFU and Important Notice. Complete and return the vigilance response section of the form to technicalservikces@randox.com.

Products

Magnesium on RX Imola analyser IVD
Brand
Randox Laboratories, Limited Ardmore 55 The Diamond Road Crumlin United Kingdom
UPC
reagent: MG3880 analyser: RX4900

Units Affected

15