Response BandLoc 5.5/6.0 Tensioner Instrument The Tensioner Instrument (01-1102-1000) is utilized to allow the surgeon to apply a recommended tension to the band and hold it in place during spine manipulation and set screw tightening.
Hazard
Over-tensioning of the band resulting in damage to the band causing it to rupture
What You Should Do
Remedy
The firm, OrthoPediatrics Corp., contacted each distributor/consignee via phone on 4/24/2017, regarding the recall. The phone call informed Consignees of issue and actions to be taken. The consignees were instructed to return of devices in their control were issued; cease further distribution of the product and quarantine it. The firm state that product will be distributed after Rework to new revision. If you have any questions, contact V.P. Regulatory Affairs, at (574) 268-6379 or email: mfox@orthopediatrics.com.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Response BandLoc 5.5/6.0 Tensioner Instrument
The Tensioner Instrument (01-1102-1000) is utilized to allow the surgeon to apply a recommended tension to the band and hold it in place during spine man
- Brand
- OrthoPediatrics Corp
- UPC
- All lots Product Number: 01-1102-1000
Units Affected
82