FDA Device

SDMS (SenTec Digital Monitoring System)

Hazard

Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).

What You Should Do

Remedy

Distributors were sent via e-mail a SenTec 'Mandatory initiated field correction action" letter dated September 15, 2016 . The letter provided a "Description of the issue, Affected products, Identify and localize affected products, Removal from use, Replacement of affected products at end customer & distributor / sub-distributor level". It also provided "action information for sub-distributors and affiliates, and end customers". For questions contact SenTec Inc. by e-mail (customersevice.usa@sentec.com) or phone (636-343-0000).

Products

SDMS (SenTec Digital Monitoring System)
Brand
SenTec AG Ringstrasse 39 Therwil Switzerland
UPC
SN 306497 to SN 306571.

Units Affected

75 (54 US; 21 =OUS)