FDA Device

Roche/Hitachi MODULAR Analytics Combination Systems Modular D/P: Clinical Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use. Modular E170: Immunoassay Analyzer Product Usage: The Roche COBAS INTEGRA 800/800 CTS Analyzer is fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids.

Hazard

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

What You Should Do

Remedy

On 7/15/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, an updated cleaning procedure, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this UMDC. On 10/5/2016, URGENT MEDICAL DEVICE CORRECTION UPDATE notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The notification updated the 7/15/16 notification to include additional affected systems and to provide updated cleaning instructions for all systems via the enclosed Operators Manual addendum.

Products

Roche/Hitachi MODULAR Analytics Combination Systems Modular D/P: Clinical Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use. Modular E170: Immunoassay Analyzer Product Usage: The
Brand
Roche Diagnostics Operations, Inc.
UPC
Not applicable

Units Affected

9014 in total