Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.

Hazard

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

What You Should Do

Remedy

DeRoyal sent a "URGENT VOLUNTARY RECALL DEROYAL HOT/COLD THERAPY UNIT" letter dated September 17, 2013. The letter identified the product, problem, and actions to be taken by the customers.

Products

Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.
Brand
DeRoyal Industries Inc
UPC
Lot Numbers: 31480713, 31480721, 31968743, 31971619, 31990844, 32005711, 32018491, 32037764, 32118020, 32710832, 32736223, 33053264, 33105221, 33141176, 33164204, 33164407, 33276273, 33329911

Units Affected

738 units