Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0
Hazard
Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.
What You Should Do
Remedy
Medical Device Correction Letters dated 10/23/13 were e-mailed to the user accounts on 10/23/13.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0
- Brand
- RAYSEARCH LABORATORIES AB Sveavaegen 9 Stockholm Sweden
- UPC
- Software version and build numbers 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4.
Units Affected
195 units domestically