Oxoid Legionella BCYE Growth Supplement SR0110A, packaged in 100 ml vials, 10/pkg. A growth supplement for the isolation of Legionella.
Hazard
Recalled product may not provide adequate recovery of Legionella pneumophila when used as directed and may result in false negative identification of Legionella pneumophila.
What You Should Do
Remedy
Thermo Fisher sent an Urgent Medical Device Recall letter dated October 17, 2014, to all affected customers.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Oxoid Legionella BCYE Growth Supplement SR0110A, packaged in 100 ml vials, 10/pkg.
A growth supplement for the isolation of Legionella.
- Brand
- Remel Inc
- UPC
- Lot Number: 1451506 Expiration: 26Feb2016
Units Affected
16/10/100 ml, vial, pkgs