Agee-WristJack Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. The non-perishable sterile system is packaged in an inner then outer tray and shipped in a cardboard sh

Hazard

A number of Agee WristJack Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.

What You Should Do

Remedy

Hand Biomechanics Lab, Inc. sent a Voluntary Medical Device Recall letter to all affected customers beginning on September 14, 2016 and sent by phone/fax numbers - or e-mail. Letters request return of devices which have not yet been used. Customers with questions were instructed to contact Customer Service at (800) 522-5778. For questions regarding this recall call 916-923-5073.

Products

Agee-WristJack Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. T
Brand
Hand Biomechanics Lab Inc
UPC
Lot number: HBL-115-165, serial numbers 47460959 and 47460980; Lot number: HBL-116-014 - all serial numbers; Lot number : HBL-116-040 - 15 systems shipped prior to recall

Units Affected

65 units