Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro diagnostic machine component.

Hazard

Possible for sample tubes in the loader to be opened and contaminate the system as well as adjacent samples during processing.

What You Should Do

Remedy

Roche Diagnostics Operations, Inc. sent an Urgent Medical Device Correction letter to affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were asked to retain the Urgent Medical Device Correction letter until a Roche Field Service Representative has installed the BLM modification kit. Consignees were also asked to complete the attached form and fax to 1-877-906-8982, ext 2015. Consignees with questions were instructed to call 1-800-428-2336. For questions regarding this recall call 317-521-3911.

Products

Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro diagnostic machine component.
Brand
Roche Diagnostics Operations, Inc.
UPC
During transport from the Bulk Loader Module (BLM), catalog number 07135645001, to the cobas p 512/p 612 pre-analytical instruments, sample tubes may inadvertently open and spill, causing contamination of the system.

Units Affected

1