HemoCue Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. Affected lots identified initially were 110705, 110706, 110715, 110716, 110717, 110718, 110719, and 110720. Subsequent investigation also identified as affected lots 1310227, 1310252
Hazard
Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are for In Vitro Diag
What You Should Do
Remedy
The consignees will be notified by a Field Safety Notice. Distributors will be instructed to forward the Field Safety Notice to their end customers and to secure that verification forms will be filled-out by end customers and returned to HemoCue.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
HemoCue Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the
- Brand
- Radiometer America Inc
- UPC
- Model Number 110705: Lot numbers: 1310500, 1310501, 1310502, 1310503, 1310518, 1310519, 1310520, 1310521, 1310767, 1310768, 1310769, 1310770, 1310771, 1310772, 1310779, 1310780, 1310781, 1310788, 1310789, 1310790, 1310799, 1311524, 1311525, 1311526, 1311527, 1311528, 1311529, 1311530, 1311531, 1311532, 1311533, 1311534, 1311535, 1311536, 1311537, 1311538, 1311539, 1311541, 1311542 Model Number 110706: Lot numbers: 1310207, 1310208, 1310209, 1310210, 1310211, 1310212, 1310213, 1310214, 1310215, 1310216, 1310228, 1310231, 1310232, 1310234, 1310235, 1310238, 1310239, 1310247, 1310249, 1310250, 1310256, 1310257, 1310258, 1310259, 1310260, 1311270, 1311271, 1311272, 1311273, 1311274, 1311279, 1311280, 1311281, 1311282, 1311283, 1311284, 1311291, 1311296, 1311297, 1311314, 1311315, 1311316, 1311317, 1311318, Model Number 110715: Lot numbers: 1310217, 1310218, 1310219, 1310221, 1310222, 1310229, 1310230, 1310237, 1310240, 1310241, 1310242, 1310251, 1310252, 1311267, 1311268, 1311269, 1311298, 1311301, 1311302, 1311303, 1311304, 1311321, 1311322, 1311323, 1311324 Model Number 110716: Lot numbers: 1310202, 1310203, 1310204, 1310223, 1310224, 1310225, 1310227, 1310243, 1310244, 1310245, 1310246, 1310261, 1311263, 1311264, 1311265, 1311266, 1311277, 1311278, 1311305, 1311306, 1311307, 1311308, 1311309, 1311310, 1311311, 1311312, 1311313, 1311325 Model Number 110717: Lot numbers: 1310504, 1310505, 1310506, 1310507, 1310508, 1310509, 1310510, 1310514, 1310515, 1310516, 1310517, 1310773, 1310774, 1310775, 1310776, 1310783, 1310784, 1310785, 1310791, 1310792, 1310794, 1310795, 1310796, 1311540, 1311543, 1311544, 1311545 Model Number 110718: Lot numbers: 1310511, 1310512, 1310513, 1310777, 1310778, 1310786, 1310787, 1310797, 1310798 Model Number 110719: Lot numbers: 1310236, 1310237, 1310248, 1311285, 1311286, 1311287, 1311288, 1311289, 1311290, 1311292, 1311293, 1311294, 1311295, 1311299, 1311300 Model Number 110720: Lot numbers: 1310205, 1310206, 1310226, 1310253, 1310254, 1310255, 1311319, 1311320
Units Affected
3,712,900 microcuvettes