FDA Device

iovera 155 Smart Tip; Catalog numbers:STT0412-10 and STT0412-5 (10 and 5 count respectively) iovera Smart Tips are individually packaged in Tyvek PET pouches, which are then packaged into 5 or 10 pack boxes and shipper boxes. Product Usage: Cyrogenic Surgical Device The myoscience iovera device is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for

Hazard

The expiration date on the outer box label and the pouch label for the iovera Smart Tip product is incorrect. Instead of indicating the correct expiration or 2015-12. the labels were incorrectly labeled 2016-05.

What You Should Do

Remedy

Myoscience sent an Urgent Medical Device Correction letters dated July 1, 2015 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers are requested to quarantine any affected product at their site. Myoscience will make arrangements to either correct the incorrect expiation date label or remove the product and replace it with correctly labeled product immediately. Question or concerns may be directed to Customers Service at 510-933-1500.

Products

iovera 155 Smart Tip; Catalog numbers:STT0412-10 and STT0412-5 (10 and 5 count respectively) iovera Smart Tips are individually packaged in Tyvek PET pouches, which are then packaged into 5 or 10 pa
Brand
Myoscience Inc
UPC
Catalog numbers:STT0412-¿10 and STT0412-¿5;  Lot numbers : 1504021-¿01 and 1505008.

Units Affected

12 boxes - 3 boxes of 10 count and 9 boxes of 5 count.