FDA Device

DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4) REF 28-0217 Insufflation Tubing, Laparoscopic Rx Only 5) REF 28-0208 Insufflation Tubing w/Vent Rx Only 6) REF 28-0212H Heated Hi-Flow Insufflation Tubing Rx Only. Used during laparoscopic surgical procedures to aid in the insufflation of the abdomen.

Hazard

DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.

What You Should Do

Remedy

The firm notified customers via letters dated 7/15/2013 sent UPS and via DeRoyal Representatives. Customers were asked to notify DeRoyal of any remaining inventory to return for replacement product or credit. Non-responders will receive additional notifications. Effectiveness checks will be performed by contacting non-responders by phone.

Products

DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4)
Brand
DeRoyal Industries Inc
UPC
All Lots of the Insufflation Tubing Sets are as follows: REF 28-0206 Insufflation Tubing, Laparoscopic, Lot #29074214; REF 28-0214 Hi-Flow Insufflation Tubing Lot #3194170; REF 28-0216 Tubing, Insuf w/.3 Micr Filter Lot #29074249; REF 28-0217 Insufflation Tubing, Laparoscopic Lot #32305010; DeRoyal REF 28-0208 Insufflation Tubing w/Vent Lot #29314302; DeRoyal REF 28-0212H Heated Hi-Flow Insufflation Tubing Lot #29000354.

Units Affected

506,632 sets in US; 17,854 sets International