FDA Device

Bigliani/Flatow The Complete Solution Fukuda Retractors Large Nonsterile 4305-19-40. Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.

Hazard

This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.

What You Should Do

Remedy

In 2010, Zimmer initiated a recall of the Bigliani/Flatow Fukuda Retractors due to fractures occurring at the grooves around the perimeter of the blades. Zimmer is expanding this recall to include seven lots that were not previously included. On 5/23/13, an Urgent Medical Device Recall notification letter went to all consignees explaining the recall expansion and risk associated with the recall. All consignees are to locate and quarantine the device immediately. Please return the product along with the completed inventory certification form to Zimmer. If you have any questions regarding this recall, call Zimmer at 877-946-2761.

Products

Bigliani/Flatow The Complete Solution Fukuda Retractors Large Nonsterile 4305-19-40. Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.
Brand
Zimmer, Inc.
UPC
Item Number: 00-4305-019-40 with Lot Numbers 61588143 and 61612276. 2010 Market Withdrawal: Item #:430501940 Lot #'s: 60768814, 60806551, 60893241, 60946820, 60999665, 61031668, 61110808, 61170678, 61205673, 61244194, 61392456, 61430127

Units Affected

722 units (includes 2010 market withdrawal)