FDA Device

Xoran MiniCAT CT Scanner ID ENT00001

Hazard

Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.

What You Should Do

Remedy

The service report in response to this complaint stated that the Head select relay board, chip and fan were replaced. The machine (ENTOOOO I) was subsequently tested. All tests were successful.

Products

Xoran MiniCAT CT Scanner ID ENT00001
Brand
Xoran Technologies, Inc.
UPC
minicat

Units Affected

320