Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer.

Hazard

A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener is supplied with the Roche COBAS INTEGRA 400/400 plus Analyzers. The Oracle TNS-Listener can be considered as an authentication and redirection component between the client and the database inside the Roche product. This means an attacker could register their own non-Roche database via remote access

What You Should Do

Remedy

Roche sent an Urgent Medical Device Correction letter on November 19, 2012, to all affected customers via UPS Ground (receipt required). The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete the attached form and return by fax and file this Medical Device Correction for future reference. For technical support customers were instructed to contact Diagnostics Technical Support at 1-900-428-2336. For questions regarding this recall call 317-521-3911.

Products

Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer.
Brand
Roche Diagnostics Operations, Inc.
UPC
Part Numbers: 28065047001 and 03245233001

Units Affected

845 Units