Dimension Calcium Flex Reagent is used on the Dimension clinical chemistry system as an in vitro diagnostic test intended for the quantitive determination of calcium activity in human serum, plasma and urine. Dimension(R) CA Flex(R) reagent is used on the Dimension(R) clinical chemistry system as an in vitro diagnostic test intended for the quantitative determination of calcium activity in human serum, plasma and urine.

Hazard

Siemens Healthcare Diagnostic has confirmed a high frequency of "Abnormal assay" flags and imprecision. Test results across the assay range may be falsely depressed or elevated, as much as, 1.93 mg/dL at a concentration of 9.26 mg/dL.

What You Should Do

Remedy

An Urgent Field Safety Notice datey July 2011 was issued to all affected customers. Customers are being advised to immediately discontinue use and discard any remaining inventory of Dimension(R) CA Flex(R0 reagent cartridge - lot EA2153. Replacement order for the affected lot is in process, and will be sent to impacted customers free of charge. Customers were instructed to forward the Field Safety Notice to anyone whom they may have distributed the affected product. For questions regarding this recall call 302-631-6299.

Products

Dimension Calcium Flex Reagent is used on the Dimension clinical chemistry system as an in vitro diagnostic test intended for the quantitive determination of calcium activity in human serum, plasma an
Brand
Siemens Healthcare Diagnostics, Inc.
UPC
Catalog # DF23A; Lot # EA2153

Units Affected

2500 cartons