Cloth electrodes with Carbon construction: P/N 201-1231, Cat Nos.: 201-123, 202-454, 203-118, 203-342, 203-454, 203-493, 203-672, 400-877, and 400-877-BJC, 2" x 2" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1241, Cat Nos.: 201-124, 202-455, 203-340, 203-674, and 400-872, 2" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1251, Cat Nos.: 201-138, 202-276, 202-456, 203-336, 203-680,

Hazard

Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.

What You Should Do

Remedy

Rich-Mar sent "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letters to affected customers beginning on 11/30/2009. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 423-648-7730 ext. 214 for questions relating to t his recall.

Products

Cloth electrodes with Carbon construction: P/N 201-1231, Cat Nos.: 201-123, 202-454, 203-118, 203-342, 203-454, 203-493, 203-672, 400-877, and 400-877-BJC, 2" x 2" Electrode, Cloth, All Se
Brand
NAImco Inc dba Richmar Inc
UPC
All Serial Numbers distributed between 11/01/2009 and 01/31/2010

Units Affected

209860 electrodes