Zoll pedi.padz Reduced Energy Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with Zoll M Series and E series Defibrillators specifically for pediatric patients.

Hazard

Lack of an adequate silicone seal around the periphery of the attenuator enclosure may allow fluid to enter.

What You Should Do

Remedy

Zoll Medical issued Urgent Device Correction notices dated 12/30/11 via UPS/Registered Mail. The letter described the issue and identified the affected product. ZOLL recommended that customers alert their staff of the issue, that the electrodes should be kept away from a fluid environment, to use caution in handling the electrodes if they are subjected to a fluid environment and maintain safe defibrillation practice, and seek a replacement if available. Also customers should contact Zoll 1-800-348-9011 for product replacement. The firm plans to replace the affected lots free of charge.

Products

Zoll pedi.padz Reduced Energy Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with Zoll M Series and E series Defibrillators specifically for pediatric patients.
Brand
Bio-Detek, Inc.
UPC
Lot Numbers 0510 thru 4511

Units Affected

3,080 units