FDA Device

Custom surgical kits and trays: GEO-MED TOTAL KNEE PACK PGYBK, REF 89-6664.04 GEO-MED TOTAL HIP PACK, REF 89-6664.03 surgical kits

Hazard

Surgical kits and trays contain Stryker Hytrel Togas which were recalled because the clear tape, that is intended to aid in the prevention of patient's fluids from contacting the surgeon/surgical staff through the sleeve seam of the toga, was not applied during the manufacturing process.

What You Should Do

Remedy

DeRoyal initiated the recall on 6/16/2012 by notifying customers by overnight delivery of letters describing the problem, instructing the customer how to handle inventory on hand, and requesting the customer to destroy the product, fill out the provided notice of destruction form, and return the form to DeRoyal. For questions customers were instructed to call 865-362-1020 or 865-362-1034. For questions regarding this recall call 865-362-2334.

Products

Custom surgical kits and trays: GEO-MED TOTAL KNEE PACK PGYBK, REF 89-6664.04 GEO-MED TOTAL HIP PACK, REF 89-6664.03 surgical kits
Brand
DeRoyal Industries Inc
UPC
GEO-MED TOTAL KNEE PACK PGYBK, REF 89-6664.04; Lot numbers: 22737521, 22924079, 23427293, 23536123, 24076552, 24435084, 24435084, 24750543, 25088133, 25275511, 25598563, 26476576, 26924324, 26959148, 27264233, 27526176, 27756781, 27756781, 28238361, 28481694   GEO-MED TOTAL HIP PACK, REF 89-6664.03; Lot numbers: 22924204, 23623163, 23880637, 24076691, 24162522, 24295421, 24585728, 24600836, 24900118, 25030484, 25144351, 25461743, 25559311, 26021993, 26064301, 27526328, 28262740

Units Affected

303 units