FDA Device

Baxter Three-Way Large Bore Stopcock with Rotating Male Luer Lock Adapter; a sterile, nonpyrogenic fluid pathway; Baxter Healthcare Corporation, Deerfield, IL 60015; product code 2C6201. Intended use: For the administration of drugs and solutions to a patient's vascular system through a vascular access device. The stopcock was sold alone and as a component of the following I.V. Administration Sets: a) product code 1C8471, Baxter Interlink System Large Bore 3-Way Stopcock with Rotating Male Lu

Hazard

Rotating the handle on the 3-Way Large Bore Stopcock past the stop can cause the stopcock to leak. A leak can lead to under-infusion or interruption of critical life-sustaining therapy.

What You Should Do

Remedy

Safety Alert letters dated October 28, 2010 were mailed via first class mail to the direct accounts, to the attention of the Director of Nursing and the Director of Materials Management, on the same date, informing them that rotating the handle on the 3-Way Large Bore Stopcock past the stop can cause the stopcock to leak. A leak can lead to under-infusion or interruption of critical life-sustaining therapy. The accounts were requested to post the enclosed attachment, which illustrates the correct direction to turn the handle of the stopcock, where the stopcocks/sets are dispensed. The accounts were also requested to complete the attached customer reply form confirming their receipt of the letter and fax it to Baxter at (847) 270-5457. Dealers, wholesalers and distributors were requested to notify their customers of this action so they can comply with the procedures listed above. The Baxter Field Corrective Action Gatekeepers were also notified of this action via e-mail on 10/27/10, and instructed to notify their customers and ministries of health of the recall. Any questions regarding the communication were directed to The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.

Products

Baxter Three-Way Large Bore Stopcock with Rotating Male Luer Lock Adapter; a sterile, nonpyrogenic fluid pathway; Baxter Healthcare Corporation, Deerfield, IL 60015; product code 2C6201. Intended use
Brand
Baxter Healthcare Corp. Rt.
UPC
all lots of product codes 2C6201, 1C8471, 2N5600, 3C0114, 3C0180, 3C0021, 3C0174, 3C0098, 3C0090 and 3C0175