Modular Microplasty Cup inserter 3/8" thread , REF 31-400600
Hazard
There was a clerical error in order entry which resulted in the recalling firm recieving spring components from a supplier made of "music wire" The recalling frim conducted tests to determine the effects of theis change in materials. The testing showed that the wire would oxidize and there would be early failure of the device.
What You Should Do
Remedy
The firm send URGENT MEDICAL DEVICE RECALL NOTICES to its distributors on 7/28/2010. The notices instructed the distributors to discontinue use of the product and return it to Biomet.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Modular Microplasty Cup inserter 3/8" thread , REF 31-400600
- Brand
- Biomet, Inc.
- UPC
- 658180, 658190 and 907630.
Units Affected
331 total for both products