Siemens Uroskop Access. Fluoroscopic, image intensified x-ray system
Hazard
Siemens has discovered a possible hazard to patients or hospital personnel when tilting the Uroskop Access system starting with serial number 6076. When tilting the table on the Uroskop Access table, the motor may become disconnected and the entire table assembly may fall on the floor.
What You Should Do
Remedy
Siemens sent a Update instructions letter dated April 2012, to all affected customers. The safety instruction letter identified the product the problem and the action needed to be taken by the customer. The update has been completed successfully. The customer has been informed of the benefits of improvements resulting from this update. Further questions please call (610) 219-4834.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Siemens Uroskop Access.
Fluoroscopic, image intensified x-ray system
- Brand
- Siemens Medical Solutions USA, Inc
- UPC
- Model number 5756130 -- serial numbers 6076, 6078, 6080, 6081, and 6082.
Units Affected
5