Philips Medical Systems, Cardiac Viewer or Pulmonary Viewer Application that can run on all Gemini TFs, GXLs and EBW Workstations. Philips Medical Systems, Cleveland, OH 44143. (The product is a software application that is installed in Philips Medical System CT scanners.)
Hazard
Incorrect Measurement: Incorrect measurement will occur when the operator uses the distance or area measurement function, such as Graphic Tool line or ROI on zoomed and batch saved images from the Cardiac Viewer or Pulmonary Viewer and then views them in the CT Viewer or on a PACS System.
What You Should Do
Consignees were notified of this recall by an Urgent Device Notification letter sent on 9/10/07. The letter instructed the users to not save any zoomed images to a batch file. Make all measurements of zoomed images in the CT viewer, which provides correct measurements. Do not perform measurements on any zoomed batched images that have been stored on a PACS System. Phillips in investigating the problem and will provide updates free of charge when completed.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Philips Medical Systems (Cleveland) Inc
- UPC
- For 16 slice, 16 slice Big Bore, 40 slice, and 64 slice Brilliance CT scanners the affected software application versions are: 2.0, 2.0.2, 2.2, 2.2.1, 2.2.2, and 2.2.5. For Extended Brilliance Workspace the affected software application versions are: 2.0.0, 2.0.1, 2.0.50, 2.1.0, 3.0.0, 3.0.1, 3.0.2, 3.5, and 3.5.2.
Units Affected
2,544