Abacus TPN Calculating Software Product Usage: LIMS Windows-based program for order entry and TPN calculation.

Hazard

Ordering salt-based ingredients on an ion-based template resulted in medication errors when calculating amounts of ingredients required.

What You Should Do

Remedy

Baxa Corporation sent a "Safety Alert" letter dated January 19, 2009 via FedEx to all affected customers. The letter described the affected product, problem, and actions to be taken. Customers were instructed to ensure that users create either a salt-based template OR an ion-based template for ordering, and that both methods of ordering are not used in one template. Customers were provided a list of safe practices to follow on the enclosed ISMP Safety Alert. Customers were advised upon receipt of the Safety Alert that Baxa Technical Support will review and address any issues with the customer on their configuration setup for calcium and magnesium. For questions contact Baxa Technical Support at 1-800-678-2292.

Products

Abacus TPN Calculating Software Product Usage: LIMS Windows-based program for order entry and TPN calculation.
Brand
Baxa Corporation
UPC
Catalogue Numbers:  8300-0045: Abacus Calculator Only (Abacus CE) 8300-0046: Abacus Single Work Station (Abacus SE) 8300-0047: Abacus Multi-Work Station (Abacus ME)

Units Affected

667 units