Ellex Motorised Safety Filter, Zeiss; Model number 6339585. Intended for photocoagulation.
Hazard
Potential for two screws important in the design of the safety filter to come loose.
What You Should Do
Ellex issued a bulletin stating the problem and identifying the affected products. Customers are to remove the Motorised Safety Filter from clinical use if the green laser light is visible while viewing through the microscope at any time during use of the Solitaire laser. The Bulletin also states that customers should contact an Ellex approved Service Engineer to investigate and perform the Field Corrective Action listed. The manufacturer will remedy the defect or bring the product into compliance free of charge. Ellex will inspect and rework all affected safety filters sold prior to February 2009.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Ellex Medical Pty Ltd
- UPC
- Solitaire LP4532 photocoagulator.
Units Affected
200 units total