Elecsys 2010 Analyzer Software version 06-07 The Elecsys 2010 analyzer is intended for the in vitro quantitative and qualitative analysis of analytes in body fluids.

Hazard

Elecsys 2010 customers have reported getting no value and flag 48 results below expected value range after performing automatic dilution for a sample that previously recovered above the measuring range. Incorrect but believable results may occur for the CEA, Rubella IgG or SHBG assays. The other affected assays will generate either no value or a negative result, both of which are not bel

What You Should Do

Remedy

Roche sent an Urgent Medical Device Correction letter dated January 31, 2012, via UPS Ground (receipt required) to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to do the following: use one of the workarounds outlined in this Urgent Medical Device Correction (UMDC) if you are running an automatic dilution on an Elecsys 2010 analyzer with software version 06-07; complete the attached faxback form (4600-00-0112) and fax it to 1-888-912-8460; and file this UMDC for future reference. For questions regarding this recall call 1-800-428-2336.

Products

Elecsys 2010 Analyzer Software version 06-07 The Elecsys 2010 analyzer is intended for the in vitro quantitative and qualitative analysis of analytes in body fluids.
Brand
Roche Diagnostics Operations, Inc.
UPC
Elecsys 2010 Analyzer Rack: 11804014001 Elecsys 2010 Analyzer Disk: 11568248001 Software version 06-07: 06526080001, 06569668001

Units Affected

Software version 06-07 was installed on Elecsys 2010 analyzers at 13 accounts.